Remote control commands
R&S
®
RTO6
1401
User Manual 1801.6687.02 ─ 05
HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST:
HIGH
Usage:
Asynchronous command
TRIGger<m>:TIMeout:TIME
<Time>
Defines the time limit for the timeout at which the instrument triggers.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Time>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
100E-9
Default unit: s
Usage:
Asynchronous command
24.9.8
Interval trigger
....................................................................................... 1401
.......................................................................................1402
........................................................................................1402
........................................................................................ 1402
TRIGger<m>:INTerval:SLOPe
<Slope>
Sets the edge for the trigger. You can analyze the interval between positive edges or
between negative edges.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope>
POSitive | NEGative
See
Chapter 24.4.3, "Slope parameter"
*RST:
POSitive
Usage:
Asynchronous command
Trigger