![R&S RTO6 User Manual Download Page 779](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646779.webp)
Protocol analysis
R&S
®
RTO6
779
User Manual 1801.6687.02 ─ 05
Reverse attenuation
Sets the attenuation factor. It is used to de-amplify the reverse signal before subtract-
ing it from the forward signal.
The effective signal amplitude passed to the decoder is:
Remote command:
Polarity
Selects the polarity of the data signal. You can select between "Normal" and "Inverted".
For "Normal" polarity, active high is defined as +1 V and active low is defined as -1 V.
For "Inverted" polarity, active high is defined as -1 V and active low is defined as +1 V.
Remote command:
Thresholds
Sets the positive and negative thresholds for the digitalization of the signal line.
The signal can have three states:
●
High: the signal value is higher than the positive threshold.
●
Low: the signal value is below the negative threshold.
●
Zero: signal is between the positive and the negative threshold.
"Thresholds"
Enter the value directly.
"Find thresh-
olds"
Executes the measurement of reference levels and sets the thresh-
olds to the middle reference level of the measured amplitude.
Remote command:
Operation mode
Selects the desired direction of the full-duplex signal for analysis. If set to "Auto", the
decoder automatically detects the "Master" or "Slave" mode.
Remote command:
Show thresholds
If enabled, the threshold levels are displayed in the diagram.
Remote command:
13.18.2.2
Advanced settings
Access: [Bus] > "Advanced" tab.
Ethernet 100BASE-T1 (option R&S
RTO6-K560)