Remote control commands
R&S
®
RTO6
1813
User Manual 1801.6687.02 ─ 05
Return values:
<AckValue>
.
The values below
–
range, increment and reset
–
are decimal
values.
Range:
0 to 1
*RST:
0
Usage:
Query only
BUS<m>:CAN:FRAMe<n>:BITRate?
Returns the primary bit rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n>
*
Selects the frame.
Return values:
<PrimaryBitRate>
Range:
0 to 100000000000
Increment: 1
*RST:
0
Default unit: bps
Usage:
Query only
BUS<m>:CAN:FRAMe<n>:CSValue?
Returns the CRC sequence value of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n>
*
Selects the frame.
Return values:
<ChecksumValue>
.
The values below
–
range, increment and reset
–
are decimal
values.
Range:
0 to 2097151
*RST:
0
Usage:
Query only
BUS<m>:CAN:FRAMe<n>:DLCValue?
Returns the data length code of the selected frame - the number of data bytes in the
frame.
Protocols