Remote control commands
R&S
®
RTO6
1294
User Manual 1801.6687.02 ─ 05
FORMat:BPATtern
<BitPatternFormat>
Sets the number format for remote bit pattern queries on serial protocols.
Parameters:
<BitPatternFormat>
DEC | HEX | OCT | BIN | ASCII | ASCii | STRG
ASCII = ASCii
*RST:
HEX
Usage:
Asynchronous command
SYSTem:DISPlay:UPDate
<Enable>
Defines whether the display is updated while the instrument is in the remote state. If
the display is switched off, the normal GUI is replaced by a static image while the
instrument is in the remote state. Switching off the display can speed up the measure-
ment. This is the recommended state.
See also:
Chapter 22.5.3.2, "Using the display during remote control"
Parameters:
<Enable>
ON | 1:
Display is shown and updated during remote control
OFF | 0:
Display shows static image during remote control
Example:
SYSTem:DISPlay:UPDate 1
Switch on the display update.
SYSTem:KLOCk
<Enable>
Locks or unlocks the local controls of the instrument. This includes the front panel
keys, the keyboard, or other local interfaces. except for the "View" button on the dis-
play.
Parameters:
<Enable>
ON | 1:
Locks the local keys
OFF | 0:
Keys are unlocked
Usage:
SCPI confirmed
GPIB:ADDRess
<Address>
Sets the GPIB address of the instrument. Changing the address has major effects on
the communication to the remote computer.
Parameters:
<Address>
Range:
0 to 30
Increment: 1
*RST:
20
Usage:
Asynchronous command
General remote settings