ZVC:BANDwidth
<Bandwidth>
Sets the bandwidth limit of the probe. The bandwidth specifies the maximum frequency
at which a purely sinusoidal signal is still transferred at 89 % (0.1 dB) of its amplitude.
Parameters:
<Bandwidth>
Range:
5000 to 1E+6
Increment: 5000
*RST:
1E+6
Default unit: Hz
Usage:
Asynchronous command
ZVC:TYPE
<DecimationMode>
Sets the decimation mode for the R&S
RT-ZVC probe. Decimation reduces the data
stream of the ADC to a stream of waveform points with lower sample rate and a less
precise time resolution.
Parameters:
<DecimationMode>
SAMPle | PDETect | HRESolution
*RST:
SAMPle
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:BANDwidth?
Queries the bandwidth of the current channel. You can set the probe bandwidth with
.
The bandwidth of some current channels is restricted to 300KHz due to their vertical
settings.
Suffix:
<m>
.
1..2
<n>
1..4
Return values:
<Bandwidth>
Range:
5000 to 1E+6
Increment: 5000
*RST:
1E+6
Default unit: Hz
Acquisition and setup