SEARch:TRIGger:WINDow:TIMerange
<SearchName>,<TimeRangeMode>
SEARch:TRIGger:WINDow:TIMerange?
<SearchName>
Selects how the time limit of the window is defined. Time conditioning is available for
the vertical conditions "WITHin" and "OUTSide" (see
Parameters:
<TimeRangeMode>
WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time
Width - Delta
and for
Width + Delta
at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than
Width - Delta
or longer than
Width +
Delta
.
SHORter
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
*RST:
WITHin
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:WINDow:WIDTh
<SearchName>,<Width>
SEARch:TRIGger:WINDow:WIDTh?
<SearchName>
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "
±
Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
.
Parameters:
<Width>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
5E-9
Default unit: s