Protocol analysis
R&S
®
RTO6
690
User Manual 1801.6687.02 ─ 05
Remote command:
BUS<m>:MILStd:MAXResponse:BITS
BUS<m>:MILStd:MAXResponse:SELect
TRIGger<m>:MILStd:MAXResponse:BITS
TRIGger<m>:MILStd:MAXResponse:SELect
Show thresholds
If enabled, the threshold levels are displayed in the diagram.
Remote command:
13.12.2.2
Display settings
Access: [Bus] > "Display" tab.
For common display settings, see
MIL Specific Label Settings
Label lists are protocol-specific. A MIL-STD-1553 label file contains four values for
each identifier:
●
"RTA": hexadecimal remote terminal address value
●
"Sub Addr": hexadecimal sub address value
●
"Sub Address Label Name": the label name corresponding to the value of the sub-
address.
●
"Symbolic label": symbolic name of addressed device, specifying the device func-
tion, and the label of the sub address.
Example: MIL PTT file
# -----------------------------------------------------------
# Labels for MIL.1553 protocol
# Column order: RT address, RT label, Subaddress, Subaddress Label
# ---------------------------------------------------------------------
@PROTOCOL_NAME = mil1553
0Ah,Engine,01h,Thrust
03h,Main panel,07h,Altimeter
03h,Main panel,01h,Speed
0Eh,Only RTA
13.12.2.3
Configuring MIL-STD-1553
For configuration assign the line to the input channel, set the threshold and the timing
conditions.
For details on configuration settings, see
Chapter 13.12.2.1, "MIL-STD-1553 configura-
1. Press the [Bus] key on the front panel.
MIL-1553 (option R&S
RTO6-K530)