Measurements
R&S
®
RTO6
315
User Manual 1801.6687.02 ─ 05
See also:
and
Hysteresis
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Remote command:
REFLevel<m>:RELative:HYSTeresis
Top outer
Defines an area above the high signal level which is still considered to be high level.
Remote command:
MEASurement<m>:REFLevel:RESult:TOUTer?
Top inner
Defines an area beneath the high signal level which is still considered to be high level.
Remote command:
MEASurement<m>:REFLevel:RESult:TINNer?
Bottom inner
Defines an area above the low signal level which is still considered to be low level.
Remote command:
MEASurement<m>:REFLevel:RESult:BINNer?
Bottom outer
Defines an area beneath the low signal level which is still considered to be low level.
Remote command:
MEASurement<m>:REFLevel:RESult:BOUTer?
Automatic measurements