Remote control commands
R&S
®
RTO6
2424
User Manual 1801.6687.02 ─ 05
..........................................................................................2429
..............................................................................................2429
BUS<m>:PARallel:DECTable<n>:SHOW
........................................................................ 2430
BUS<m>:PARallel:DECTable<n>:COUNt?
......................................................................2430
BUS<m>:PARallel:DECTable<n>:DATA?
........................................................................ 2430
BUS<m>:PARallel:STATe
<Enable>
Enables or disables the indicated parallel bus. The threshold settings of the bus take
effect for all
active
parallel buses.
Dependencies: At least one digital channel must be enabled for the selected bus, oth-
erwise the command does not work. The bus is enabled automatically if the first digital
channel is enabled with
BUS<m>:PARallel:BIT<n>[:STATe]
or
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
Parameters:
<Enable>
ON | OFF
*RST:
OFF
BUS<m>:PARallel:BIT<n>[:STATe]
<Assigned>
Assigns the selected digital channel to the indicated bus, displays it, and enables the
bus if the bus was disabled. That is,
are set to ON automatically.
If another active bus already uses the selected digital channel, the instrument disables
the other bus to avoid conflicts.
For parallel bus 1, the
BUS:PAR:BIT[:STATe]
command has the same effect as
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
<n>
0..15
Selects the bit of the bus word. Each bit corresponds to a digital
channel.
Parameters:
<Assigned>
ON | OFF
*RST:
OFF
BUS<m>:PARallel:THReshold<n>
<Threshold>
Sets the logical threshold for the indicated channel group.
Mixed signal option (MSO, R&S
RTO6-B1)