DIGital<m>:DISPlay
<Display>
Enables or disables the indicated digital channel, displays it, and enables the parallel
Bus1 if the bus was disabled. That is,
are set to ON automatically.
If another active bus already uses the selected digital channel, the instrument disables
the other bus to avoid conflicts.
For Bus1, the
DIG:DISP
command has the same effect as
. To enable digital channels for buses 2, 3 and 4, use the
BUS:PAR:BIT[:STAT]
command.
Suffix:
<m>
.
0..15
Number of the digital channel
Parameters:
<Display>
ON | OFF
DIGital<m>:TECHnology
<Technology>
Selects the threshold voltage for various types of integrated circuits and applies it to all
digital channels.
The setting affects only the settings of the first MSO bus (Bus1). You can set the tech-
nology value for all buses with
.
Suffix:
<m>
.
0..15
The suffix is irrelevant.
Parameters:
<Technology>
V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
See
DIGital<m>:THReshold
<Value>
Sets the logical threshold for the channel group to which the indicated digital channel
belongs. The setting affects only the settings of the first MSO bus (Bus1).
The setting affects only the settings of the first MSO bus (Bus1). You can set the
threshold for all buses with
.
See also:
Mixed signal option (MSO, R&S
RTO6-B1)