Remote control commands
R&S
®
RTO6
2425
User Manual 1801.6687.02 ─ 05
Alternatively, you can set the threshold with
the parallel bus 1, you can also use
.
See also:
Suffix:
<m>
.
1..4
Selects the parallel bus.
<n>
1..4
Selects the channel group:
1 = dig. channels 0..3;
2 = dig. channels 4..7
3 = dig. channels 8..11
4 = dig. channels 12..15
Parameters:
<Threshold>
Range:
-8.0 to 8.0
Increment: 200.0e-12
*RST:
0
Default unit: V
BUS<m>:PARallel:TECHnology
<Technology>
Selects the threshold voltage for various types of integrated circuits and applies it to all
digital channels.
Suffix:
<m>
.
1..4
Selects the parallel bus.
Parameters:
<Technology>
V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15: TTL
V25: CMOS 5.0 V
V165: CMOS 3.3 V
V125: CMOS 2.5 V
V09: CMOS 1.85 V
VM13: ECL, -1.3 V
V38: PECL
V20: LVPECL
V0: Ground
MANual: Set a user-defined threshold value with
*RST:
V165
BUS<m>:PARallel:THCoupling
<LevelCoupling>
Sets the threshold for all digital channels of the selected bus to the same value. Also
the hysteresis value is applied to all digital channels.
Tor parallel bus 1, the command
has the same effect.
Mixed signal option (MSO, R&S
RTO6-B1)