Remote control commands
R&S
®
RTO6
1421
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Mode>
TIME | EVENts | RANDom | AUTO | OFF
TIME
Defines the holdoff directly as a time period. The next trigger
occurs only after the "Holdoff time" has passed (defined using
EVENts
Defines the holdoff as a number of trigger events. The next trig-
ger occurs only when this number of events is reached. The
number of triggers to be skipped is defined using
RANDom
Defines the holdoff as a random time limited by
on page 1423. For each acquisition cycle, the instrument selects
a new random holdoff time from the specified range.
AUTO
The holdoff time is calculated automatically based on the current
horizontal scale.
OFF
No holdoff
*RST:
OFF
Usage:
Asynchronous command
TRIGger<m>:HOLDoff:TIME
<Time>
Defines the holdoff time period. The next trigger occurs only after this time has passed.
The setting is relevant if the holdoff mode is set to TIME.
See also:
●
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Time>
Range:
100E-9 to 10
Increment: 200E-6
*RST:
1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE TIME
TRIGger:HOLDoff:TIME 1ms
The holdoff time is set to 1 ms.
Trigger