Remote control commands
R&S
®
RTO6
2408
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<ReferenceSource> CLOCk | CDR | PROTocol
*RST:
CLOCk
Usage:
Asynchronous command
EYE<m>:MEASurement:MAXPrecision
<EnabClassifier>
If this function is enabled, the rising and falling transitions for the eye measurmens
"Eye rise time", "Eye fall time" and "Duty cycle distortion" are separated.
Suffix:
<m>
.
1..4
Parameters:
<EnabClassifier>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Display
EYE<m>:MSLices
<MaxSliceCount>
Sets the number of waveform slices for a single acquisition.
Suffix:
<m>
.
1..4
Parameters:
<MaxSliceCount>
Range:
1 to 1000000
Increment: 1
*RST:
2000
Usage:
Asynchronous command
EYE<m>:DISPlay:MODE
<EyestripeMode>
Defines when the eye stripe is shown: always, never, or during mask tests.
Suffix:
<m>
.
1..4
Parameters:
<EyestripeMode>
ALWays | MTONly | NEVer
MTONly: Mask test only
*RST:
MTONly
Usage:
Asynchronous command
Protocols