Parameters:
<HystEnable>
Range:
-25 to 25
Increment: 0.1
*RST:
0.1
Default unit: V
BUS<m>:CMSB:NRZ:THReshold:PRESet
<PresetValue>
Sets the NRZ thresholds to predefined or individually definable voltage levels.
Suffix:
<m>
.
1..4
Parameters:
<PresetValue>
V05 | V2 | V5 | V7 | MANual
V05
Sets the clock and data threshold to +0.5
V (in case of NRZ
Unclocked: data threshold, only)
V2
Sets the clock and data threshold to +2.0
V (in case of NRZ
Unclocked: data threshold, only)
V5
Sets the clock and data threshold to +5.0
V (in case of NRZ
Unclocked: data threshold, only)
V7
Sets the clock and data threshold to +7.0
V (in case of NRZ
Unclocked: data threshold, only)
MANual
Allows to set individual threshold voltage levels
*RST:
V5
BUS<m>:CMSB:NRZ:THReshold:COUPling
<ThresholdCoupling>
Couples the clock and data threshold values for the NRZ Clocked coding standard.
The values are coupled to the same number.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdCoupling> ON | OFF
ON
Activates coupling of the NRZ clock and data threshold values.
OFF
Disables coupling of the NRZ clock and data threshold values.
*RST:
ON