Protocol analysis
R&S
®
RTO6
775
User Manual 1801.6687.02 ─ 05
Remote commands:
●
●
SEARch:RESult:RFFE:SEQuence<m>:TYPE?
●
SEARch:RESult:RFFE:SEQuence<m>:STATe?
●
SEARch:RESult:RFFE:SEQuence<m>:STARt?
●
SEARch:RESult:RFFE:SEQuence<m>:STOP?
●
SEARch:RESult:RFFE:SEQuence<m>:SADD?
●
SEARch:RESult:RFFE:SEQuence<m>:ADDRess?
●
SEARch:RESult:RFFE:SEQuence<m>:BCOunt?
●
SEARch:RESult:RFFE:SEQuence<m>:DATA?
●
SEARch:RESult:RFFE:SEQuence<m>:SYMBol?
●
SEARch:RESult:RFFE:SEQuence<m>:BYTE<n>:STATe?
●
SEARch:RESult:RFFE:SEQuence<m>:BYTE<n>:VALue?
13.17.6.3
Searching RFFE data
Prerequisites: An RFFE bus is configured, see
Chapter 13.17.2, "RFFE configuration"
on page 757, and "Decode" is enabled.
1. Press the [Bus] key and selet the "Shortcuts" tab.
2. Tap "Setup search".
3. If the dialog box does not contain a search entry, tap the "Add" icon to create one,
as described in
"To create a user-defined search"
4. Tap the "Source" button and select the serial bus that is set to RFFE.
"Protocol" shows the RFFE icon.
5. Tap "Type" and select the search type.
All trigger types are also available for search.
6. To refine the search settings, configure additional settings, which are available for
many search types.
For details, see
Chapter 13.17.6.1, "RFFE search setup"
7. Under "Control", "Enable" the search.
The "Search Results" box opens.
8. Close the "Search" dialog box.
9. Press [Run Stop] to start acquisition.
10. To start searching the acquired waveform for specific events, tap "Enable" in the
search setup dialog:
The R&S
RTO6 displays the "Search Results" box that lists the detected events.
RFFE (option R&S
RTO6-K550)