![R&S RTO6 User Manual Download Page 421](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646421.webp)
Search functions
R&S
®
RTO6
421
User Manual 1801.6687.02 ─ 05
11.2.1
Search criteria
Access: "Menu" > "Apps" > "Analysis" tab > "Search" > "Setup" tab
Enable
If you enable a search and run an acquisition, the search is performed continuously on
the acquired data until acquisition is stopped.
If acquisition is stopped and you enable a search, the data of the last acquisition is
searched.
Remote command:
Source
Defines the waveform to be searched. The source can be any analog and digital input
signal, math or reference waveform, or track. While the instrument triggers only on real
input signals, it can search also calculated and restored waveforms.
If the source is an FFT spectrum, you can perform a frequency marker search by using
the cursor measurement and defining the peak excursion. See
For some serial protocol options, search on a serial bus is available. For details, see
the relevant chapters of the "Protocol Analysis" chapter.
Search setup