Remote control commands
R&S
®
RTO6
1277
User Manual 1801.6687.02 ─ 05
SEARch:RESult:SENT:FCOunt? 'Search1'
SEARch:RESult:SENT:FRAMe1:STAT? 'Search1'
SEARch:RESult:SENT:FRAMe1:IDValue? 'Search1'
SEARch:RESult:SENT:FRAMe1:NIBB1:VALue? 'Search1'
SEARch:RESult:SENT:FRAMe1:NIBB2:VALue? 'Search1'
SEARch:RESult:SENT:FRAMe1:CSValue? 'Search1'
SEARch:TRIGger:SENT:ENHanced 'Search1', OFF
24.3.8
Power analysis (option R&S
RTO6-K31)
24.3.8.1
Auto deskew
Configures the voltage and current probes for power measurements and executes the
auto deskew.
Command description in
If the instrument refuses to accept
POWer
commands, activate the power mode
using
:POWer:ENABle
.
In the following example
*OPC
prevents overlapping execution of asynchronous com-
mands.
*RST; *OPC?
STOP;*OPC?
//Activate two channels
:CHANnel1:STATe 1
:CHANnel2:STATe 1
//Activate power
:POWer:ENABle
//Select current and voltage sources
:POWer:SOURce:VOLTage1 CHANnel1
:POWer:SOURce:CURRent1 CHANnel2
*OPC?
//Configure voltage probe on CH1 manually
//Not necessary if you use an active R&S voltage probe that is recognized by the instrument
:PROBe1:SETup:ATTenuation:MODE Manual
:PROBe:SETup:ATTenuation:DEFProbe ZD01a100
*OPC?
//selected high voltage differential probe 1:100
Programming examples