Remote control commands
R&S
®
RTO6
2520
User Manual 1801.6687.02 ─ 05
24.23.1
General
.................................................................................. 2520
...................................................................................2520
POWer:ENABle
Activates the power mode and initializes the power measurements. If the power mode
is disabled, the instrument does not accept any
POWer
command.
Use
POWer:ENABle
after each
*RST
.
Example:
See
Chapter 24.3.8.1, "Auto deskew"
Usage:
Event
Asynchronous command
POWer:SOURce:CURRent<1..2>
<CurrentSource>
Sets the channel for the current source.
Parameters:
<CurrentSource>
CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
*RST:
CURRent1: CHAN2, CURRent2: CHAN4,
Usage:
Asynchronous command
POWer:SOURce:VOLTage<1..4>
<VoltageSource>
Sets the channel for the voltage source input.
Parameters:
<VoltageSource>
CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
*RST:
VOLTage1: CHAN1, VOLTage2: CHAN3, VOLT-
age3: CHAN3, VOLTage4: CHAN4
Usage:
Asynchronous command
24.23.2
Deskew
Chapter 24.3.8.1, "Auto deskew"
Power analysis (option R&S
RTO6-K31)