Remote control commands
R&S
®
RTO6
2108
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:HBTO:CRC:MAX
<SearchName>,<CRCPatternMax>
SEARch:TRIGger:HBTO:CRC:MAX?
<SearchName>
Sets the end value of the frame check range if
is set to
INRange
or
OORange
.
Parameters:
<CRCPatternMax>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:DCONdition
<SearchName>,<DataOperator>
SEARch:TRIGger:HBTO:DATA:DCONdition?
<SearchName>
Sets the condition for the data. You can define an exact value or a value range.
Parameters:
<DataOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:DMIN
<SearchName>,<DataMin>
SEARch:TRIGger:HBTO:DATA:DMIN?
<SearchName>
Specifies the data bit pattern, or sets the start value of a pattern range.
Parameters:
<DataMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols