Remote control commands
R&S
®
RTO6
2107
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:HBTO:LENGth:MAX
<SearchName>,<TypePatternMax>
SEARch:TRIGger:HBTO:LENGth:MAX?
<SearchName>
Sets the end value of the length / type range if
is set to
INRange
or
OORange
.
Parameters:
<TypePatternMax>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:CRC:CONDition
<SearchName>,<CRCOperator>
SEARch:TRIGger:HBTO:CRC:CONDition?
<SearchName>
Sets the condition for the frame check. You can define an exact value or a value range.
Parameters:
<CRCOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:CRC:MIN
<SearchName>,<CRCPatternMin>
SEARch:TRIGger:HBTO:CRC:MIN?
<SearchName>
Specifies the frame check bit pattern, or sets the start value of a pattern range.
Parameters:
<CRCPatternMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols