Remote control commands
R&S
®
RTO6
1385
User Manual 1801.6687.02 ─ 05
Parameters:
<Offset>
Range:
-100E-9 to 100E-9
Increment: 10E-15
*RST:
0
Default unit: s
Usage:
Asynchronous command
PROBe<m>:SKEState
<ProbeSkew>
If enabled, the skew of all connected active probes is measured, displayed, and used
for deskewing.
Suffix:
<m>
.
1..4
The suffix is irrelevant. The setting affects all active channels.
Parameters:
<ProbeSkew>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
CHANnel<m>:SKEW:EXTended
<ExtendSkew>
Enables a higher skew value range that can be set with
on page 1384. The skew extension reduces the maximum number of acquisitions in
the memory.
Suffix:
<m>
.
1..4
The suffix is irrelevant. The setting affects all active channels.
Parameters:
<ExtendSkew>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
24.8.12
AUX OUT
................................................................................. 1386
..........................................................................................1386
Acquisition and setup