Remote control commands
R&S
®
RTO6
1603
User Manual 1801.6687.02 ─ 05
MTESt:EYEMask:HPOSition
<MaskTestName>,<PatOffsetX>
MTESt:EYEMask:HPOSition?
<MaskTestName>
Sets the horizontal (time) value of the eye shape enter and thus defines the horizontal
position of the eye shape on the display.
Parameters:
<PatOffsetX>
Range:
-100 to 100
Increment: 0.01
*RST:
0
Default unit: s
Parameters for setting and query:
<MaskTestName>
String with the name of the mask test
Usage:
Asynchronous command
MTESt:EYEMask:VPOSition
<MaskTestName>,<PatOffsetY>
MTESt:EYEMask:VPOSition?
<MaskTestName>
Sets the vertical (voltage) value of the eye shape enter and thus defines the vertical
position of the eye shape on the display.
Parameters:
<PatOffsetY>
Range:
-100 to 100
Increment: 0.01
*RST:
0
Default unit: V
Parameters for setting and query:
<MaskTestName>
String with the name of the mask test
Usage:
Asynchronous command
24.15
Search
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands
*OPC
,
*OPC?
or
*WAI
can be used after the command or a command set.
For more information, see:
●
www.rohde-schwarz.com/rc-via-scpi
, chapter "Command Sequence and Synchro-
nization"
●
Chapter 24.3, "Programming examples"
Search