Remote control commands
R&S
®
RTO6
1373
User Manual 1801.6687.02 ─ 05
ZVC:Z<m>:V<n>:BANDwidth?
Queries the bandwidth of the voltage channel. You can set the probe bandwidth with
.
Suffix:
<m>
.
1..2
<n>
1..4
Return values:
<Bandwidth>
Range:
5000 to 1E+6
Increment: 5000
*RST:
1E+6
Default unit: Hz
Usage:
Query only
Asynchronous command
ZVC:Z<m>:V<n>:IMPedance
<MeasImp>
Sets the voltage impedance of the probe channel for power calculations and measure-
ments.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<MeasImp>
Range:
1 to 100E+3
Increment: 1
*RST:
50
Default unit: Ω
Usage:
Asynchronous command
ZVC:Z<m>:V<n>:OFFSet
<VerticalOffset>
Sets the vertical offset for the voltage channel.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<VerticalOffset>
Range:
-1 to 1
Increment: 0.01
*RST:
0
Default unit: V
Usage:
Asynchronous command
Acquisition and setup