Remote control commands
R&S
®
RTO6
2223
User Manual 1801.6687.02 ─ 05
<n>
*
Specifies the error number.
Parameters:
<Enable>
ON | OFF
*RST:
ON
Example:
TRIGger:USBPd:ERRor1:ENABle ON
Enables CRC Error.
TRIGger:USBPd:ERENable "Length Error",ON
Enables Length Error.
TRIGger<m>:USBPd:FRENable
<Frame>, <Enabler>
TRIGger<m>:USBPd:FRAMe<n>:ENABle
<Enable>
Enables or disables the checking condition for the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n>
*
Specifies the frame number.
Parameters:
<Enable>
ON | OFF
*RST:
OFF
Example:
TRIGger:USBPd:FRAMe1:ENABle ON
Enables searching for for the Test frame.
TRIGger:USBPd:FRENable "Control Frame" ON
Enables searching for the Control frame.
TRIGger<m>:USBPd:FIENable
<Frame>, <Field>, <Enabler>
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:ENABle
<CondEnabler>
Enables or disables the checking condition for a specific data pattern in the selected
field of the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n>
*
Specifies the frame.
<o>
*
Specifies the field number within the frame.
Parameters:
<CondEnabler>
ON | OFF
ON
Checking condition enabled
Protocols