Remote control commands
R&S
®
RTO6
1360
User Manual 1801.6687.02 ─ 05
Parameters:
<ProbeBandwidth>
Range:
10000 to 16E+9
Increment: 10
*RST:
4E+9
Default unit: Hz
Usage:
Asynchronous command
PROBe<m>:DEEMbedding:MODE
<Mode>
Sets the probing mode.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<Mode>
PROBing | TERMinating
PROBing
For measurements with a high-ohmic probe and measure on an
existing line in parallel to the load.
TERMinating
For measurements if the measuring equipment is the load of the
line.
*RST:
PROBing
Usage:
Asynchronous command
PROBe<m>:DEEMbedding:REMProbeload
<RemoveProbeLoad>
Removes the loading of the probe if
DEEMbedding<m>:COMPonent<n>:MODE
is set
to
PROBing
.
If the probe loading is removed, you see the signal that would be at the measurement
point if the probe's input impedance is ideal (infinite impedance).
Without removing the loading, you see the real signal at the measurement point,
including the probe loading.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<RemoveProbeLoad> ON | OFF
*RST:
ON
Usage:
Asynchronous command
Acquisition and setup