Remote control commands
R&S
®
RTO6
1347
User Manual 1801.6687.02 ─ 05
24.8.7.2
Micro button and R&S ProbeMeter
........................................................................................... 1347
..................................................................................... 1348
....................................................................................... 1348
PROBe<m>:PMETer:RESults:SINGle?
...........................................................................1349
PROBe<m>:PMETer:RESults:COMMon?
....................................................................... 1349
PROBe<m>:PMETer:RESults:DIFFerential?
................................................................... 1349
PROBe<m>:PMETer:RESults:NEGative?
....................................................................... 1350
PROBe<m>:PMETer:RESults:POSitive?
........................................................................ 1350
PROBe<m>:SETup:MODE
<Mode>
Select the action that is started when you press the micro button on the probe head.
See also:
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Mode>
RCONtinuous | RSINgle | AUToset | AZERo | SEToffsettomean |
SITFile | NOACtion | FINDtriglevel | REPort | PROBemode |
PRSetup
RCONtinuous
Run continuous: The acquisition is running as long as the probe
button is pressed.
RSINgle
Run single: Starts a defined number of acquisitions (same as
[Single] key).
AUTOSET
Starts the autoset procedure.
AZero
AutoZero: performs an automatic correction of the zero error.
SEToffsettomean
Set offset to mean: performs an automatic compensation for a
DC component of the input signal.
SITFile
Save Image To File:
Directs the display image to a file. The
com-
mand defines the file name. The file format is defined with
NOACtion
Nothing is started on pressing the micro button.
FINDtriglevel
Sets the trigger level automatically to
0.5 * (MaxPeak – Min-
Peak)
. The function is not available for an external trigger
source.