Remote control commands
R&S
®
RTO6
2464
User Manual 1801.6687.02 ─ 05
24.20
Pulse Source and TDR/TDT Analysis (options
R&S
RTO6-B7/K130)
24.20.1
Pulse source (option R&S
RTO6-B7)
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands
*OPC
,
*OPC?
or
*WAI
can be used after the command or a command set.
For more information, see:
●
www.rohde-schwarz.com/rc-via-scpi
, chapter "Command Sequence and Synchro-
nization"
●
Chapter 24.3, "Programming examples"
PSRC[:STAT]
<State>
Switches the pulse output on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
PSRC:OUTPutlow
<OutputLowLevel>
Sets the low level of the output pulse.
Parameters:
<OutputLowLevel>
Range:
-0.2 to -0.05
Increment: 0.01
*RST:
-0.05
Default unit: V
Usage:
Asynchronous command
Pulse Source and TDR/TDT Analysis (options R&S
RTO6-B7/K130)