Protocol analysis
R&S
®
RTO6
831
User Manual 1801.6687.02 ─ 05
SEU
Defines, whether an SEU check that meets specific conditions is executed or not. S
and E represent the Start and End of a start-split transaction, U represents the
reserved/Unused bit of a complete-split transaction. Permissible binary SEU values are
00, 01, 10, and 11.
"Condition"
Defining a specific SEU value or an SEU value range requires to set
the operator to one of the following: equal, not equal, less than, less
than or equal, greater than, greater than or equal, in range, out of
range.
"Min"
Specifies an SEU value, or sets the start value of an SEU value
range.
"Max"
Sets the end value of an SEU value range if "Condition"
(
INRange
or
OORange
.
Remote command:
ET
Defines, whether an Endpoint Type (ET) check that meets specific conditions is execu-
ted or not.
"Condition"
Defining a specific endpoint type or an ET range requires to set the
operator to one of the following: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies an ET, or sets the start value of an ET range.
"Max"
Sets the end value of an ET range if "Condition" (
INRange
or
OORange
.
Remote command:
13.20.3.2
Triggering on USB 2.0
Prerequisite: A bus is configured for the USB 2.0 signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus] key and selet the "Shortcuts" tab.
2. Press "Setup trigger".
USB (option R&S
RTO6-K570)