Remote control commands
R&S
®
RTO6
1626
User Manual 1801.6687.02 ─ 05
WIDTh
Defines a time range for keeping up the true result of the pattern
condition. The range is defined using
*RST:
OFF
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
SEARch:TRIGger:PATTern:TIMeout:MODE
<SearchName>,<TimeoutMode>
SEARch:TRIGger:PATTern:TIMeout:MODE?
<SearchName>
Defines the condition for the timeout.
Parameters:
<TimeoutMode>
HIGH | LOW | EITHer
EITHer
High or low, the pattern remains unchanged for the given time.
*RST:
HIGH
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
SEARch:TRIGger:PATTern:TIMeout[:TIME]
<SearchName>,<Time>
SEARch:TRIGger:PATTern:TIMeout[:TIME]?
<SearchName>
Defines how long the result of the pattern condition must keep the given state.
Parameters:
<Time>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
100E-9
Default unit: s
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
SEARch:TRIGger:PATTern:WIDTh:RANGe
<SearchName>,<WidthRangeMode>
SEARch:TRIGger:PATTern:WIDTh:RANGe?
<SearchName>
Defines the time range of a pulse width for keeping up the true result of the pattern
condition. The width and delta are specified using
SEARch:TRIGger:PATTern:WIDTh:DELTa
Parameters:
<WidthRangeMode> WITHin | OUTSide | SHORter | LONGer
Search