![R&S RTO6 User Manual Download Page 1535](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461535.webp)
Remote control commands
R&S
®
RTO6
1535
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
MEASurement<m>:STATistics:HBINs
<HistogramBins>
Sets the number of bins - the number of vertical bars that build the histogram.
If
is ON, the instrument determines the number
of bins automatically based on the time base, the current measurements, and other
settings.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<HistogramBins>
Range:
2 to 1000
Increment: 10
*RST:
1000
Usage:
Asynchronous command
MEASurement<m>:STATistics:MODE
<ResetMode>
Defines when the statistics for long term measurements are reset.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<ResetMode>
TIME | WFMS | MEAS
TIME
Sets one long term measurement point after the time defined
using
MEASurement<m>:STATistics:RTIMe
WFMS - Waveforms
Sets one long term measurement point after a number of
acquired waveforms defined using
MEAS
Sets one long term measurement point after a number of mea-
surement results.
*RST:
TIME
Usage:
Asynchronous command
MEASurement<m>:STATistics:RCOunt
<RstWfmCnt>
Defines the number of measured waveforms from which one point of the long term
measurement is created (reset of statistics).