Remote control commands
R&S
®
RTO6
1536
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<RstWfmCnt>
Range:
1 to 65535
Increment: 10
*RST:
1000
Usage:
Asynchronous command
MEASurement<m>:STATistics:RMEascount
<RstMeasCnt>
Defines the number of measurement results from which one point of the long term
measurement is created.
This setting is only available if
MEASurement<m>:STATistics:MODE
MEAS
.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<RstMeasCnt>
Range:
1 to 65535
Increment: 10
*RST:
1000
Usage:
Asynchronous command
MEASurement<m>:STATistics:RTIMe
<ResetTime>
Defines the time or period after which the statistics are reset.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<ResetTime>
Range:
0.1 to 2.9
Increment: 1E-3
*RST:
0.2
Default unit: s
Usage:
Asynchronous command
MEASurement<m>:STATistics:RESet
Resets the histogram, the long term measurement and the statistics.