Remote control commands
R&S
®
RTO6
1538
User Manual 1801.6687.02 ─ 05
MEASurement<m>:VERTical:SCALe
<VerticalScale>
Defines the vertical scaling per division, so that the scaling can be adapted automati-
cally during the long term measurement period.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<VerticalScale>
Range:
1E-12 to 100E+12
Increment: 10E-6
*RST:
0.5
Default unit: V/div
Usage:
Asynchronous command
MEASurement<m>:LTMeas[:STATe]
<ShowDiagram>
Enables long term measurement for a defined number of measurement points (see
on page 1538) or a specified time
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<ShowDiagram>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
MEASurement<m>:LTMeas:COUNt
<MeasCnt>
Defines the total number of points to be measured during the long term measurement.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<MeasCnt>
Range:
1000 to 200000
Increment: 10
*RST:
1000
Usage:
Asynchronous command
Automatic measurements