Remote control commands
R&S
®
RTO6
1539
User Manual 1801.6687.02 ─ 05
MEASurement<m>:LTMeas:TIME
<MeasurementTime>
Defines the total duration of the long term measurement.
This setting is only available if
MEASurement<m>:STATistics:MODE
"Time".
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<MeasurementTime> Range:
0.01 to 2.9
Increment: 1
*RST:
200
Default unit: s
Usage:
Asynchronous command
24.12.11
Protocol
Before starting the protocol measurement, a bus type has to be selected (
) and the bus has to be activated (
The following table lists the <MeasType> parameter values with a short description.
Table 24-11: Protocol measurements
<MeasType> param-
eter value (remote
control)
Measurement (man-
ual operation)
Description/Result
AUDio
Audio
FLDValue
Field value
Value of a field over time
F2F
Frame to frame
The distance between two frame types
Trigger to frame
GAP
Gap
Measures a gap, periods at which the bus is idle. The dis-
tance of a gap can only be measured between two identi-
fied frames.
MBITrate
Main bit rate
Bit rate of the protocol
SBITrate
2nd bit rate
Additional bit rate, protocol dependent.
BIDLe
Bus idle
Calculates the bus idle time
FCNT
Frame count
Number of all frames within the acquisition window
FEC
Frame error count
Sum of all frames with errors within the acquisition window
Automatic measurements