Command Reference
R&S
®
ZNB/ZNBT
736
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<FreqPowTime>
Frequency, power or time value, to be defined in accordance
with the sweep type (
Range:
Virtually no restriction for limit segments.
*RST:
A segment that is created implicitly, e.g. by means
of CALCulate<Ch>:LIMit:UPPer[:DATA] or CALCu-
late<Ch>:LIMit:LOWer[:DATA,], covers the maxi-
mum sweep range of the analyzer.
Default unit: Hz for frequency sweeps, dBm for power sweeps, s
for time sweeps
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
CALC:LIM:SEGM:STIM:STAR 1GHZ; STOP 2 GHZ; :
CALC:LIM:SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a stimu-
lus range between 1 GHz and 2 GHz.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1000000000,2000000000,2,3.
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE
<LimLineType>
Selects the limit line type for a limit line segment. This can be done before or after
defining the stimulus and response values of the segment, however, a segment must
be created first to enable this command (e.g
CALC:LIM:DATA
).
Note:
The type command overwrites the
and is overwritten by them. It is not affected by the other commands in the
CALCulate<Chn>:LIMit...
subsystem defining stimulus and response values of
limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
SCPI Command Reference