Command Reference
R&S
®
ZNB/ZNBT
1000
User Manual 1173.9163.02 ─ 38
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
Prepares the analyzer for fixture compensation comprising a single or a series of fix-
ture compensation sweeps (
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
ACQuire]
). Previous compensation data is deleted.
Suffix:
<Ch>
.
Channel number of the corrected channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected
<CalGroupFile>,
<Standard>, <TestPort1>[, <SecondPortOrAdapter>]
Reloads a set of previously acquired calibration data for a particular standard from a
file in the cal pool. The loaded data may be combined with new calibration measure-
ment data (
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
) in order
to simplify and speed up the new calibration procedure. The channel settings for loa-
ded and new calibrations (e.g. the number of sweep points) must be identical.
Note:
●
The analyzer performs a consistency check for the loaded data. If the loaded file is
incompatible with the channel settings of channel
<Ch>
, or if it does not contain
data for the specified standard and port(s), a command error message (
–
100,
"Command error;...") is generated.
●
For a sliding match, the R&S
ZNB/ZNBT can acquire and load measurement data
for up to 20 positions per port. Multiple calls of
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected <CalGroupFile>,
SLIDe, <TestPort>
for the same cal group file and test port, implicitly increases
the position until position 20 has been recorded. Subsequent calls will start over at
position 1, overwriting the previously loaded data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalGroupFile>
String parameter to specify the name of the loaded cal group
file.
SCPI Command Reference