Command Reference
R&S
®
ZNB/ZNBT
1186
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<UnitRef>
S | SINV | Y | Z | YREL | ZREL
Keyword for the physical unit of the response values. The
parameters form four groups: S and SINV select relative units
(dB) for the limit line. Y selects admittance units (S/Siemens). Z
selects impedance units (Ω). YREL and ZREL select dimension-
less numbers (U).
Usage:
Setting only
CALCulate<Chn>:LIMit:RDOMain:FORMat
<UnitRef>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via
.
Tip:
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:SPACing
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<UnitRef>
COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values. The
parameters form four groups: COMPlex, REAL, IMAGinary, and
SWR select dimensionless numbers (U) for the limit line. MAG-
Nitude selects relative units (dB). PHASe selects phase units
(deg). GDELay selects time units (s). L selects inductance units
(H/Henry). C selects capacitance units (F/Farad).
Usage:
Setting only
CALCulate<Chn>:LIMit:RDOMain:SPACing
<Format>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via
.
Tip:
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:FORMat
Suffix:
<Chn>
.
Channel number used to identify the active trace.
R&S ZVR/ZVAB Compatible Commands