
Command Reference
R&S
®
ZNB/ZNBT
790
User Manual 1173.9163.02 ─ 38
Example:
See
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
Manual operation:
See
"Renormalization according to Theory of"
CALCulate<Chn>:TRANsform:TIME[:TYPE]
<TransformType>
Selects the time domain transformation type.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TransformType>
BPASs | LPASs
BPASs - band pass impulse (only impulse response; a step
response
CALCulate<Chn>:TRANsform:TIME:STIMulus
STEP
is automatically changed to impulse response)
LPASs - low pass (impulse or step response, depending on
CALCulate<Chn>:TRANsform:TIME:STIMulus
*RST:
BPASs
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME LPAS; TIME:STIM STEP
Select a low pass step transformation.
CALC:TRAN:TIME:LPAS KFST
Calculate a harmonic grid, keeping the stop frequency and the
number of points.
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:CENTer
<CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterTime>
Center time of the diagram in time domain.
Range:
-99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
SCPI Command Reference