Command Reference
R&S
®
ZNB/ZNBT
995
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Port numbers of the analyzer
Note:
If the analyzer is set to use the same connectors at all
[SENSe<Ch>:]CORRection:COLLect:CONNection:
), then a change of a connector type is valid for all
ports. The gender of the connectors can still be different.
Parameters:
<ConnectorType>
N50Female | N50Male | N75Female | N75Male | PC7 |
SMAFemale | SMAMale | PC35female | PC35male |
PC292female | PC292male | BNC50male | BNC50female |
BNC75male | BNC75female | UFEMale1 | UMALe1 |
UFEMale2 | UMALe2
Connector type and gender of the connectors (omitted for
query). The R&S ZVR-compatible parameters
UFEMALE1
and
UMALE1
denote the user-defined connector type "UserConn1",
UFEMALE2
and
UMALE2
denote the user-defined connector type
"UserConn2". The user-defined connector types must be defined
before being addressed by
[SENSe<Ch>:]CORRection:COLLect:
CONNection<PhyPt>
.
*RST:
N50FEMALE for all ports.
Example:
*RST; :CORR:COLL:CONN1 N75MALE; CONN4?
Change the connector type at port 1 from N50FEMALE to
N75MALE. The connector type at the other ports is also
changed to N75, however, the gender (female) is maintained.
CORR:COLL:CONN4?
returns
N75FEMALE
.
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
<Gender>
Qualifies whether the genders at the test ports (but not their connector types) are equal
or independent.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<Gender>
ALL | SINGle
ALL
–
equal (uniform) genders. If the gender at one port is
changed, the connector type at all other ports is changed
accordingly.
SINGle
–
independent (possibly non-uniform) genders at the
ports.
*RST:
SINGle
SCPI Command Reference