Command Reference
R&S
®
ZNB/ZNBT
811
User Manual 1173.9163.02 ─ 38
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:DATA<Port>
<Interchange>, <arg1>
Defines an embedding network for a balanced port based on the given S-Parameter
traces.
Circuit models
STSL | STSC | SLST | SCST
require S-Parameter traces of two 2-
port networks, to be assigned to the different ports; the
FIMPort
model requires S-
Parameter traces of a single 4-port network but no additional port assignment.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<LogPt>
cuit data from a Touchstone file located at the R&S
ZNB/ZNBT's file system.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number
<Port>
Port assignment for two 2-port networks:
1
- Port 1
2
- Port 2
This parameter is ignored for 4-port networks.
Setting parameters:
<Interchange>
FPORts | IPORts
FPORts - the port numbers are not interchanged.
IPORts - Interchange port numbers of loaded trace data.
This parameter is ignored for 4-port networks.
<arg1>
<block_data>
Content of a Touchstone file (
*.s2p
or
*.s4p
) in IEEE488.2
.
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G<Cmp> in the different circuit models for balanced
port embedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
SCPI Command Reference