Command Reference
R&S
®
ZNB/ZNBT
1134
User Manual 1173.9163.02 ─ 38
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Manual operation:
See
" Settling Delay / Reset Delay "
SOURce<Ch>:POWer<PhyPt>:STARt
<StartPower>
SOURce<Ch>:POWer<PhyPt>:STOP
<StopPower>
These commands define the start and stop powers for a power sweep. The values also
define the left and right edges of a Cartesian diagram. A power sweep must be active
(
POWer
) to use these commands.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected power sweep range applies to all source ports used
in the active channel.
Parameters:
<StopPower>
Start or stop power of the sweep.
Range:
-150 dBm to +100 dBm. The usable power range is
frequency-dependent; refer to the data sheet.
Increment: 0.01 dB
*RST:
-25 dBm start power, 0 dBm stop power
Default unit: dBm
Example:
SWE:TYPE POW
Activate a power sweep.
SOUR:POW:STAR -6; STOP 10
Select a power sweep range between -6 dBm and +10 dBm.
Manual operation:
See
"Start Frequency / Stop Frequency / Power CW Frequency"
Note:
If the start power entered is greater than the current stop power
(
SOURce<Ch>:POWer<PhyPt>:STOP
), the stop power is set to the start power plus
the minimum power span (increment) of 0.01 dB.
If the stop power entered is smaller than the current start power
(
SOURce<Ch>:POWer<PhyPt>:STARt
), the start power is set to the stop power
minus the minimum power span of 0.01 dB.
SCPI Command Reference