
Annexes
R&S
®
ZNB/ZNBT
1264
User Manual 1173.9163.02 ─ 38
To
Universal
Interface
100
+5 V
22 pF
10
Figure 10-7: Circuit diagram of the output path
The default configuration of the signals is listed below. Notice that
*RST
or "Preset" do
not change the configuration of the Universal Interface. Use
to restore default values.
Signal
*RST Configuration
Port A, Port B
Output ports, all bits "High" (decimal 0)
Port C, Port D
Input ports
/PORT C STATUS, /PORT D STATUS
"Low"
/OUTPUT 1, /OUTPUT 2
"High"
/SWEEP END
"High"
/PASS_FAIL
"High"
SCPI commands:
(resets all configurable signals)
10.2.4.4
Timing of Control Signals
The timing of the essential measurement control signals is shown in the figure below.
The duration of the shaded time intervals depends on the measurement settings.
Interfaces and Connectors