GUI Reference
R&S
®
ZNB/ZNBT
502
User Manual 1173.9163.02 ─ 38
n
i
i
y
n
RMS
1
2
|
|
1
Note:
To calculate the "Min", "Max", "Pk-Pk" and the "Std Dev" values, the analyzer
uses formatted response values y
i
(see trace formats). Consequently, the mean value
and the standard deviation of a trace depend on the selected trace format. In contrast,
the "RMS" calculation is based on linear, unformatted values. The physical units for
unformatted wave quantities is 1 Volt. The RMS value has zero phase. The selected
trace format is applied to the unformatted RMS value, which means that the RMS
result of a trace does depend on the trace format.
Remote command:
CALCulate<Chn>:STATistics[:STATe]
CALCulate<Chn>:STATistics:RESult?
CALCulate<Chn>:STATistics:MMPTpeak[:STATe]
CALCulate<Chn>:STATistics:MSTDdev[:STATe]
CALCulate<Chn>:STATistics:RMS[:STATe]
Format
This determines how
Min/Max/Peak-Peak, Mean/Std Dev/RMS
traces (Smith, Polar) are calculated:
●
"ZVAB": the results are based on unformatted wave quantities (voltages)
●
"R + jX": the results are based on the impedance values R and X
●
"G + jB": the results are based on the admittance values G and B
In the two latter cases, the "RMS" value is not displayed.
Remote command:
CALCulate<Chn>:STATistics:FORMat
Phase/El Length
Displays or hides the phase delay ("Phs Dly") and the electrical length ("El Len") of the
trace in the selected evaluation range ("Eval. Range..."). The parameters are only
available for trace formats that contain phase information, i.e. for the formats "Phase" ,
"Unwr Phase" , and the polar diagram formats "Polar" , "Smith" , "Inv Smith" (see
Chapter 5.10, "Format Softtool"
on page 469). Moreover, the sweep type must be a
frequency sweep, and the evaluation range must contain at least 3 measurement
points.
The phase parameters are obtained from an approximation to the derivative of the
phase with respect to frequency in the selected evaluation range.
●
"Phs Dly" is the phase delay, which is an approximation to the group delay and cal-
culated as follows:
f
PD
deg
360