Command Reference
R&S
®
ZNB/ZNBT
1195
User Manual 1173.9163.02 ─ 38
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'A1'
Create channel 4 and a trace named "Ch4Tr1" to measure the
wave quantity a
1
. The trace automatically becomes the active
trace.
OUTP4:DPOR PORT2
Select drive port 2 for the active trace.
[SENSe<Ch>:]CORRection:CKIT:<ConnType>:<StandardType>
<CalkitName>,
<Standard>, <MinFreq>, <MaxFreq>, <DelayParam>, <Loss>, <C0>, <L0>,
<C1>, <L1>, <C2>, <L2>, <C3>, <L3>, OPEN | SHORt
Defines the parameters of a calibration standard
<StandardType>
for a specified
connector type
<ConnType>
. A particular physical standard can be selected by speci-
fying the name of the calibration kit and its serial number. Depending on the standard
type, only a subset of the parameters may be used; see
.
Note:
If the specified cal kit does not exist, it is created with the specified calibration
standard.
Suffix:
<Ch>
.
Channel number. This suffix is ignored because calibration kits
are channel-independent.
Parameters:
<ConnType>
Connector type, one of the following identifiers:
N50, N75:
N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292:
PC 7, PC 3.5 or 2.92 mm connectors
USER<no>:
User-defined connectors UserConn1, UserConn2
SMA:
User-defined connector type SMA ...
Note:
This command only supports ZVR-compatible connector
types. For general definitions use
<StandardType>
Standard type. For reflection standards, the first character
denotes the gender, e.g.:
FOPEN, MOPEN:
Open (f) or Open (m) standard.
For transmission standards, the first two characters denotes the
genders on both ends, e.g.:
FFSNetwork, MFSNetwork, MMSNetwork:
Symm. network
(ff), symm. network (mf) or symm. network (mm) standard.
For a complete list of standard types refer to
.
Parameter list
String parameters to specify the configured standard
(
<CalkitName>
,
<StandardLabel>
) and numeric parameters
defining its properties. See
.
*RST:
n/a
R&S ZVR/ZVAB Compatible Commands