SYSTem:CORRection:WIZard[:IMMediate]
<Dialogs>
Keysight-compatible command to open the
.
Tip:
When working with the
Chapter 6.1.2, "GPIB Explorer"
on page 642, switch to
raw mode ("Options">"Raw mode" in the IECWIN32 GUI) before executing this com-
mand.
Setting parameters:
<Dialogs>
MAIN | CKIT
MAIN: open the "Calibration > Start Cal" softtool tab
CKIT: open the "Calibration Kits" dialog
Usage:
Setting only
SYSTem:DATE
<Year>, <Month>, <Day>
The command queries or defines the instrument's current date setting.
The setting command requires administrator rights; refer to
Parameters:
<Year>
Year, four-digit number
<Month>
Month, two-digit number, 01 (for January) to 12 (for December)
<Day>
Day, two-digit number, 01 to the number of days in the month
Example:
SYST:DATE?
Response:
2012,05,01
- it is the 1st of May, 2012.
SYSTem:DFPRint?
Queries the device footprint. The device footprint contains detailed information about
the instrument and is mostly used for service purposes.
Usage:
Query only
Manual operation:
See
"Save... / Print... / Save Report"