Command Reference
R&S
®
ZNB/ZNBT
1170
User Manual 1173.9163.02 ─ 38
SYSTem:PRESet:USER[:STATe]
<Boolean>
Selects a factory preset or a user-defined preset.
Note:
The user-defined preset can be initiated using "System > Preset" (manual con-
trol) or
MMEMory:LOAD:STATe
.
*RST
and
SYSTem:PRESet[:DUMMy]
always initiate
a factory preset.
Parameters:
<Boolean>
OFF - user-defined preset switched off (factory preset is used).
ON - user-defined preset switched on.
Example:
SYST:PRES:USER ON
Enable a user-defined preset.
SYST:PRES:USER:NAME 'C:\Users\Public
\Documents\Rohde-Schwarz\Vna\RecallSets\Setup_2.znx'
Select a setup file for the user-defined preset.
Press PRESET (manual control) or use
MMEM:LOAD:STAT 1, 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\RecallSets\Setup_2.znx'
.
Manual operation:
See
SYSTem:PRESet:USER:CAL
<PresetUserCal>
Selects a calibration from the calibration pool that shall be restored during a user-
defined preset.
The corresponding cal group file (
<PresetUserCal>.cal
) must be available in the
cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Data\
.
Parameters:
<PresetUserCal>
Name of the cal group (i.e. the name of the cal group file without
path and extension).
Use the empty string to restore the default behavior (no preset
user cal).
Example:
MMEMORY:STORE:CORRection 1, 'Refl1
Trans123.cal'
Adds the active user calibration to the memory pool
SYSTem:PRESet:USER:CAL 'Refl1 Trans123'
Sets the newly created cal pool member as the preset user cal.
SYSTem:PRESet:USER:CAL?
Queries for the preset user cal. Returns
'Refl1 Trans123'
SYSTem:PRESet:USER:CAL ''
Restores the default behaviour of the instrument: no preset user
cal.
Manual operation:
See
SCPI Command Reference