Command Reference
R&S
®
ZNB/ZNBT
749
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number in the range 1 to 10. This numeric suffix is
ignored and may be set to any value.
Example:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPAN
Sets the sweep span of the sweep range equal to the absolute value of the first coordi-
nate of the active delta marker
<Mk>
on trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number in the range 1 to 10.
Example:
*RST; :CALC:MARK ON; MARK:DELTa ON
Create marker 1 in the center of the current sweep range and
enable the delta mode.
CALC:MARK:X 300MHz
Increase the stimulus value of the delta marker by 300 MHz.
CALC:MARK:FUNC:SPAN
Set the sweep range equal to 300 MHz. The sweep range starts
at the reference marker position, i.e. in the center of the analyz-
er's frequency range.
Usage:
Event
Manual operation:
See
"Center = Marker / Start = Marker / Stop = Marker / Span =
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
These command sets the beginning (
...STARt
) and the end (
...STOP
) of the sweep
range equal to the stimulus value of the marker
<Mk>
on trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number in the range 1 to 10
SCPI Command Reference