Command Reference
R&S
®
ZNB/ZNBT
1152
User Manual 1173.9163.02 ─ 38
<Driver>
Power meter type (string parameter). The power meter type is
identical with the name of the power meter driver file (
*.pwm
)
stored in the
resources\extdev
subdirectory of the analyz-
er's program directory. Type the driver file name as shown in the
"Add External Generator" dialog, i.e. without the file extension
"
.pwm
" (example: use '
NRVD
', if the corresponding driver file
name is
NRVD.pwm
).
Alternative:
The '<Driver>' string may also contain the power
meter driver file name with its complete path, e.g. '
C:\Program
Files\Rohde&Schwarz\Network
Analyzer\resources\extdev\nrvd.pwm
'.
<Interface>
Interface type (string parameter):
'GPIB0'
,
'VXI-11'
,
'socket'
,
'other'
,
'USB'
(for the supported USB devi-
ces) ...
<Address>
Interface address (string parameter), depending on the interface
type. See
Example:
SYST:COMM:RDEV:PMET:CONF:AUTO OFF
Disables "Auto Config NRP-Zxx" (if it was enabled previously).
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter',
'NRP-Z55', 'usb', '100045'
Configure an R&S NRP power meter as external power meter
no. 1, assigning the name "USB Power Meter" and a serial num-
ber 100045.
SYST:COMM:RDEV:PMET:DEL
Clear the power meter configuration table.
SYST:COMM:RDEV:PMET:DEF?
Query the power meter configuration. The analyzer returns an
error message because the power meter no. 1 is no longer con-
figured.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:SPCorrection[:STATe]
<Boolean>
Gets/sets the state of the built-in S-parameter correction that is available on certain
R&S®NRP-Z power sensors.
Note that this state is persistently stored on the power sensor (and NOT on the
R&S
ZNB/ZNBT).
See Application Note 1GP70 "Using S-Parameters with R&S®NRP-Z Power Sensors"
for background information. This Application Note is available on the Rohde & Schwarz
internet at
http://www.rohde-schwarz.com/appnotes/1GP70
Suffix:
<Pmtr>
.
Number of the configured power meter.
Parameters:
<Boolean>
ON | OFF
SCPI Command Reference