GUI Reference
R&S
®
ZNB/ZNBT
398
User Manual 1173.9163.02 ─ 38
Figure 5-30: Scalar Power Calibration
3. Perform a complementary source flatness calibration (optional).
Figure 5-31: Source Flatness Calibration (LO: internal second source)
During source flatness calibration the DUT should be connected with all ports
defined in the mixer setup. First the LO ports are calibrated. In the second step the
source flatness of the RF and IF ports is calibrated while the LO ports are activa-
ted.
5.5.6.5
SMARTerCal (Manual) Wizard for Mixer Mode
Allows to perfom a variant of a manual "SMARTerCal" that is specially tailored to the
"Mixer Mode".
Access:
CHANNEL – CHANNEL CONFIG > "Mixer Mode" > "SMARTerCal" – "Start...
(Manual)"
Offers similar functionality as the general
lowing differences:
●
Always all channels are calibrated
●
The affected test ports are already preselected and labeled according to their role
in the mixer measurement
●
The adequate calibration type is already preselected
Channel Config Softtool