System error correction
In general, the system error correction is no longer valid after a change of the sweep
parameters. The status of the calibration is shown in the trace list. If the number of
points is changed, the analyzer interpolates the correction data. The calibration label
"Cal Int" is displayed. See also
Chapter 4.5.4, "Calibration Labels"
Number of Points
Sets the total number of measurement points per sweep. The minimum number of
points is 1 (measurement at a single frequency/power/time value). The maximum
depends on the analyzer type.
Together with the sweep range defined in the
on page 430 of the "Stimulus" softtool, this parameter defines the grid of sweep points.
The step width between two consecutive sweep points is constant on a linear scale
(sweep types "Lin Freq", "Time" and "CW Mode") or on a logarithmic scale (sweep
types "Log Freq" and "Power").
The sweep points for linear frequency sweeps may alternatively be defined using the
.
Tip: Measurement time and screen resolution
A large number of points improves the resolution of the trace but increases the mea-
surement time. The overall measurement time is composed of the hardware settling
time at the beginning of the sweep plus the sum of the measurement times at each
individual sweep point. This implies that the measurement time increases roughly line-
arly with the number of points.
See also
Chapter 4.1.4, "Sweep Control"
Remote command:
Freq Step Size
Sets the distance between two consecutive sweep points of a linear frequency sweep.
Sweep Softtool