GUI Reference
R&S
®
ZNB/ZNBT
461
User Manual 1173.9163.02 ─ 38
5.9.8 Imbal. CMRR Tab
Selects an imbalance or common mode rejection ratio (CMRR) as measured quantity.
These measurements are available if at least one balanced port is active.
Background information
Refer to the following sections:
●
Chapter 4.3.5, "Unbalance-Balance Conversion"
●
Chapter 4.3.5.3, "Imbalance and Common Mode Rejection"
5.9.8.1
Controls on the Imbal. CMRR Tab
The Imbalance Differential... button opens the Opens a dialog of the same name that
allows you to measure the differential mode imbalance of a DUT with balanced input
and output ports, connected to a balanced test port and two single-ended test ports
Imbalance/CMRR
parameter as a measured quantity for the active trace.
These parameters are expressed as "Imb
<out><in>
" or "CMRR
<out><in>
", where <out> and
<in> denote the logical output and input port numbers of the DUT.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "IMB21" |
"IMB12" | "CMRR11" | "CMRR21" | "CMRR12" ...
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "IMB21" |
"IMB12" | "CMRR11" | "CMRR21" | "CMRR12" ...
Meas Softtool